Print This Page

Spec #72300

SUPREME TECHNOLOGIES, INC.

Supreme Power  ® 

Specification for  Sealed Rechargeable Nickel Cadmium  Battery

 Supreme Power LifeLight #72300

Specification of Battery SP
Type  NiCd
  SupremePower LifeLight  #72300
Chemical System
Nickel Cadmium
NiCd
Type Otoscope/Medical Battery Medical Use
Niminal Voltage
Standard
3.6 V
Nominal Capacity
Low Rate-0.1C
 
Capacity Charge Discharge
Minimum
Typical
Weight   95 G
   
Standard
Quick *
Fast *
Charging Minimum Charge
75
mA
75
mA 
75
mA (0.1C)
Time Required (*hrs)
16
Hrs
16
Hrs
16
Hrs
Maximum Charge 150
mA 
1300
mA
 750
mA 
Time Required (hrs)  8
Hours
150
min
72 Min's. 
(or-10m^V)
Minimum Overcharge
75
mA 
for 
24 
hours 
Maximum Overcharge
750
mA
for 
90
min 

(with Cut-Off) 
Maximum Discharge Current Continuous 2250
mA
       
Momentary
(1 second)
3.75
A
       
Internal Impedance Typical at 1000Hz
30
milliohms at 50% Discharge
    Storage for <1 Month (C)     Storage for <1 Year(C)
Temperature Minimum
-20
-10
Maximum
40
30
 
Discharge ("C)
Charge ("C)
Minimum
-10
0
Maximum
50
40
Service Life
Standard (IEC61436)
500 Cycles
cycles (for reference only)
Designations
IEC-60285
Appearance and Dimensions (mm)

Refer to CAD Drawing

Height
Diameter
H1
MAX73.5
MAX28.0
0.3 (REF)
Components No Side Tab            

* Quick and fast charge requires charge control circuitry to terminate charge or switch to trickle charge when cell reaches full charge.
Remark: The information contained herein is presented only as a guide for the applications of our products. Data in this specification are subjected to change without notice and become contractual only after written confirmation by STI (Hong Kong) Ltd.

?2005 Supreme Power Battery/Supreme Technologies. Reproduction is strictly forbidden

Specification data sheets are the sole property of Supreme Technologies, Inc., Supreme Power Battery, STI(Hong Kong)Lmt., Penny R Stepro


Previous page: Spec#72200